Q-2, r. 38 - Regulation respecting the quality of the atmosphere

Full text
67.4. (Replaced, O.C. 501-2011)Design criteria: A biomedical waste incinerator must be designed and operated in such a way as to offer at all times a gas residence time of at least 1.0 second in the final combustion chamber, at a temperature of 1 000ºC. It must be equipped with a parallel combustion system that uses gas or a type of oil defined in section 1.
O.C. 584-92, s. 4; O.C. 501-2011, s. 215.